Matching Items (20)
Filtering by

Clear all filters

152336-Thumbnail Image.png
Description
Flow measurement has always been one of the most critical processes in many industrial and clinical applications. The dynamic behavior of flow helps to define the state of a process. An industrial example would be that in an aircraft, where the rate of airflow passing the aircraft is used to

Flow measurement has always been one of the most critical processes in many industrial and clinical applications. The dynamic behavior of flow helps to define the state of a process. An industrial example would be that in an aircraft, where the rate of airflow passing the aircraft is used to determine the speed of the plane. A clinical example would be that the flow of a patient's breath which could help determine the state of the patient's lungs. This project is focused on the flow-meter that are used for airflow measurement in human lungs. In order to do these measurements, resistive-type flow-meters are commonly used in respiratory measurement systems. This method consists of passing the respiratory flow through a fluid resistive component, while measuring the resulting pressure drop, which is linearly related to volumetric flow rate. These types of flow-meters typically have a low frequency response but are adequate for most applications, including spirometry and respiration monitoring. In the case of lung parameter estimation methods, such as the Quick Obstruction Method, it becomes important to have a higher frequency response in the flow-meter so that the high frequency components in the flow are measurable. The following three types of flow-meters were: a. Capillary type b. Screen Pneumotach type c. Square Edge orifice type To measure the frequency response, a sinusoidal flow is generated with a small speaker and passed through the flow-meter that is connected to a large, rigid container. True flow is proportional to the derivative of the pressure inside the container. True flow is then compared with the measured flow, which is proportional to the pressure drop across the flow-meter. In order to do the characterization, two LabVIEW data acquisition programs have been developed, one for transducer calibration, and another one that records flow and pressure data for frequency response testing of the flow-meter. In addition, a model that explains the behavior exhibited by the flow-meter has been proposed and simulated. This model contains a fluid resistor and inductor in series. The final step in this project was to approximate the frequency response data to the developed model expressed as a transfer function.
ContributorsHu, Jianchen (Author) / Macia, Narciso (Thesis advisor) / Pollat, Scott (Committee member) / Rogers, Bradley (Committee member) / Arizona State University (Publisher)
Created2013
Description
Optical Fibers coupled to laser light sources, and Light Emitting Diodes are the two classes of technologies used for optogenetic experiments. Arizona State University's Flexible Display Center fabricates novel flexible Organic Light Emitting Diodes(OLEDs). These OLEDs have the capability of being monolithically fabricated over flexible, transparent plastic substrates and having

Optical Fibers coupled to laser light sources, and Light Emitting Diodes are the two classes of technologies used for optogenetic experiments. Arizona State University's Flexible Display Center fabricates novel flexible Organic Light Emitting Diodes(OLEDs). These OLEDs have the capability of being monolithically fabricated over flexible, transparent plastic substrates and having power efficient ways of addressing high density arrays of LEDs. This thesis critically evaluates the technology by identifying the key advantages, current limitations and experimentally assessing the technology in in-vivo and in-vitro animal models. For in-vivo testing, the emitted light from a flat OLED panel was directly used to stimulate the neo-cortex in the M1 region of transgenic mice expressing ChR2 (B6.Cg-Tg (Thy1-ChR2/EYFP) 9Gfng/J). An alternative stimulation paradigm using a collimating optical system coupled with an optical fiber was used for stimulating neurons in layer 5 of the motor cortex in the same transgenic mice. EMG activity was recorded from the contralateral vastus lateralis muscles. In vitro testing of the OLEDs was done in primary cortical neurons in culture transfected with blue light sensitive ChR2. The neurons were cultured on a microelectrode array for taking neuronal recordings.
ContributorsShah, Ankur (Author) / Muthuswamy, Jitendran (Thesis advisor) / Greger, Bradley (Committee member) / Blain Christen, Jennifer (Committee member) / Arizona State University (Publisher)
Created2015
153963-Thumbnail Image.png
Description
There is a strong medical need and important therapeutic applications for improved wireless bioelectric interfaces to the nervous system. Multichannel devices are desired for neural control of robotic prosthetics that interface to remaining nerves in limb stumps of amputees and as alternatives to traditional wired arrays used in for some

There is a strong medical need and important therapeutic applications for improved wireless bioelectric interfaces to the nervous system. Multichannel devices are desired for neural control of robotic prosthetics that interface to remaining nerves in limb stumps of amputees and as alternatives to traditional wired arrays used in for some types of brain stimulation. This present work investigates a new approach to ultrasound-powering of implantable microelectronic devices within the tissue that may better support such applications. These devices are of ultra-miniature size that is enabled by a wireless technique. This study investigates two types of ultrasound-powered neural interfaces for multichannel sensory feedback in neurostimulation. The piezoceramics lead zirconate titanate (PZT) ceramic and polyvinylidene fluoride (PVDF) polymer were the primary materials used to build the devices. They convert ultrasound to electricity that when rectified by a diode produce a current output that is neuro stimulatory to peripheral nerve or the neurons in the brain. Multichannel devices employ a form of spatial multiplexing that directs focused ultrasound towards localized and segmented regions of PVDF or PZT that allows independent channels of nerve actuation. Different frequencies of ultrasound were evaluated for best results. Firstly, a 2.25 MHz frequency signal that is reasonably penetrating through body tissue to an implant several centimeters deep and also a 5 MHz frequency more suited to application for actuation of devices within a less than a centimeter of nerve. Results show multichannel device performance to have a complex inter-relationship with frequency, size and thickness, angular incidence, channel separations, and number of folds (layers connected in series and parallel). The output electrical port impedances of PVDF devices were examined in relationship to that of stimulating electrodes and tissue interfaces. Miniature multichannel devices were constructed using an unreported method of employing state of the art laser cutting systems. The results show that PVDF based devices have advantages over PZT, because of better acoustic coupling with tissue, known better biocompatibility, and better separation between multiple channels. However, the PZT devices proved to be better overall in terms of compactness and higher outputs for a given ultrasound power level.
ContributorsNanda Kumar, Yashwanth (Author) / Towe, Bruce (Thesis advisor) / Muthuswamy, Jitendran (Committee member) / Nikkhah, Mehdi (Committee member) / Arizona State University (Publisher)
Created2015
153889-Thumbnail Image.png
Description
Robust and stable decoding of neural signals is imperative for implementing a useful neuroprosthesis capable of carrying out dexterous tasks. A nonhuman primate (NHP) was trained to perform combined flexions of the thumb, index and middle fingers in addition to individual flexions and extensions of the same digits. An array

Robust and stable decoding of neural signals is imperative for implementing a useful neuroprosthesis capable of carrying out dexterous tasks. A nonhuman primate (NHP) was trained to perform combined flexions of the thumb, index and middle fingers in addition to individual flexions and extensions of the same digits. An array of microelectrodes was implanted in the hand area of the motor cortex of the NHP and used to record action potentials during finger movements. A Support Vector Machine (SVM) was used to classify which finger movement the NHP was making based upon action potential firing rates. The effect of four feature selection techniques, Wilcoxon signed-rank test, Relative Importance, Principal Component Analysis, and Mutual Information Maximization was compared based on SVM classification performance. SVM classification was used to examine the functional parameters of (i) efficacy (ii) endurance to simulated failure and (iii) longevity of classification. The effect of using isolated-neuron and multi-unit firing rates was compared as the feature vector supplied to the SVM. The best classification performance was on post-implantation day 36, when using multi-unit firing rates the worst classification accuracy resulted from features selected with Wilcoxon signed-rank test (51.12 ± 0.65%) and the best classification accuracy resulted from Mutual Information Maximization (93.74 ± 0.32%). On this day when using single-unit firing rates, the classification accuracy from the Wilcoxon signed-rank test was 88.85 ± 0.61 % and Mutual Information Maximization was 95.60 ± 0.52% (degrees of freedom =10, level of chance =10%)
ContributorsPadmanaban, Subash (Author) / Greger, Bradley (Thesis advisor) / Santello, Marco (Thesis advisor) / Helms Tillery, Stephen (Committee member) / Arizona State University (Publisher)
Created2015
156803-Thumbnail Image.png
Description
Severe cases of congenital heart defect (CHD) require surgeries to fix the structural problem, in which artificial grafts are often used. Although outcome of surgeries has improved over the past decades, there remains to be patients who require re-operations due to graft-related complications and the growth of patients which results

Severe cases of congenital heart defect (CHD) require surgeries to fix the structural problem, in which artificial grafts are often used. Although outcome of surgeries has improved over the past decades, there remains to be patients who require re-operations due to graft-related complications and the growth of patients which results in a mismatch in size between the patient’s anatomy and the implanted graft. A graft in which cells of the patient could infiltrate, facilitating transformation of the graft to a native-like tissue, and allow the graft to grow with the patient heart would be ideal. Cardiac tissue engineering (CTE) technologies, including extracellular matrix (ECM)-based hydrogels has emerged as a promising approach for the repair of cardiac damage. However, most of the previous studies have mainly focused on treatments for ischemic heart disease and related heart failure in adults, therefore the potential of CTE for CHD treatment is underexplored. In this study, a hybrid hydrogel was developed by combining the ECM derived from cardiac tissue of pediatric CHD patients and gelatin methacrylate (GelMA). In addition, the influence of incorporating gold nanorods (GNRs) within the hybrid hydrogels was studied. The functionalities of the ECM-GelMA-GNR hydrogels as a CTE scaffold were assessed by culturing neonatal rat cardiomyocytes on the hydrogel. After 8 days of cell culture, highly organized sarcomeric alpha-actinin structures and connexin 43 expression were evident in ECM- and GNR-incorporated hydrogels compared to pristine GelMA hydrogel, indicating cell maturation and formation of cardiac tissue. The findings of this study indicate the promising potential of ECM-GelMA-GNR hybrid hydrogels as a CTE approach for CHD treatment.

As another approach to improve CHD treatment, this study sought the possibility of performing a proteomic analysis on cardiac ECM of pediatric CHD patient tissue. As the ECM play important roles in regulating cell signaling, there is an increasing interest in studying the ECM proteome and the influences caused by diseases. Proteomics on ECM is challenging due to the insoluble nature of ECM proteins which makes protein extraction and digestion difficult. In this study, as a first step to perform proteomics, optimization on sample preparation procedure was attempted.
ContributorsSugamura, Yuka (Author) / Nikkhah, Mehdi (Thesis advisor) / Smith, Barbara (Committee member) / Willis, Brigham (Committee member) / Arizona State University (Publisher)
Created2018
156937-Thumbnail Image.png
Description
In medical imaging, a wide variety of methods are used to interrogate structural and physiological differences between soft tissues. One of the most ubiquitous methods in clinical practice is Magnetic Resonance Imaging (MRI), which has the advantage of limited invasiveness, soft tissue discrimination, and adequate volumetric resolution. A myriad of

In medical imaging, a wide variety of methods are used to interrogate structural and physiological differences between soft tissues. One of the most ubiquitous methods in clinical practice is Magnetic Resonance Imaging (MRI), which has the advantage of limited invasiveness, soft tissue discrimination, and adequate volumetric resolution. A myriad of advanced MRI methods exists to investigate the microstructural, physiologic and metabolic characteristics of tissue. For example, Dynamic Contrast Enhanced (DCE) and Dynamic Susceptibility Contrast (DSC) MRI non-invasively interrogates the dynamic passage of an exogenously administered MRI contrast agent through tissue to quantify local tracer kinetic properties like blood flow, vascular permeability and tissue compartmental volume fractions. Recently, an improved understanding of the biophysical basis of DSC-MRI signals in brain tumors revealed a new approach to derive multiple quantitative biomarkers that identify intrinsic sub-voxel cellular and vascular microstructure that can be used differentiate tumor sub-types. One of these characteristic biomarkers called Transverse Relaxivity at Tracer Equilibrium (TRATE), utilizes a combination of DCE and DSC techniques to compute a steady-state metric which is particularly sensitive to cell size, density, and packing properties. This work seeks to investigate the sensitivity and potential utility of TRATE in a range of disease states including Glioblastomas, Amyotrophic Lateral Sclerosis (ALS), and Duchenne’s Muscular Dystrophy (DMD). The MRC measures of TRATE showed the most promise in mouse models of ALS where TRATE values decreased with disease progression, a finding that correlated with reductions in myofiber size and area, as quantified by immunohistochemistry. In the animal models of cancer and DMD, TRATE results were more inconclusive, due to marked heterogeneity across animals and treatment state. Overall, TRATE seems to be a promising new biomarker but still needs further methodological refinement due to its sensitivity to contrast to noise and further characterization owing to its non-specificity with respect to multiple cellular features (e.g. size, density, heterogeneity) that complicate interpretation.
ContributorsFuentes, Alberto (Author) / Quarles, Chad C (Thesis advisor) / Kodibagkar, Vikram (Thesis advisor) / Greger, Bradley (Committee member) / Arizona State University (Publisher)
Created2018
153632-Thumbnail Image.png
Description
Intracranial pressure is an important parameter to monitor, and elevated intracranial pressure can be life threatening. Elevated intracranial pressure is indicative of distress in the brain attributed by conditions such as aneurysm, traumatic brain injury, brain tumor, hydrocephalus, stroke, or meningitis.

Electrocorticography (ECoG) recordings are invaluable in understanding epilepsy and

Intracranial pressure is an important parameter to monitor, and elevated intracranial pressure can be life threatening. Elevated intracranial pressure is indicative of distress in the brain attributed by conditions such as aneurysm, traumatic brain injury, brain tumor, hydrocephalus, stroke, or meningitis.

Electrocorticography (ECoG) recordings are invaluable in understanding epilepsy and detecting seizure zones. However, ECoG electrodes cause a foreign body mass effect, swelling, and pneumocephaly, which results in elevation of intracranial pressure (ICP). Thus, the aim of this work is to design an intracranial pressure monitoring system that could augment ECoG electrodes.

A minimally invasive, low-cost epidural intracranial pressure monitoring system is developed for this purpose, using a commercial pressure transducer available for biomedical applications. The system is composed of a pressure transducer, sensing cup, electronics, and data acquisition system. The pressure transducer is a microelectromechanical system (MEMS)-based die that works on piezoresistive phenomenon with dielectric isolation for direct contact with fluids.

The developed system was bench tested and verified in an animal model to confirm the efficacy of the system for intracranial pressure monitoring. The system has a 0.1 mmHg accuracy and a 2% error for the 0-10 mmHg range, with resolution of 0.01 mmHg. This system serves as a minimally invasive (2 mm burr hole) epidural ICP monitor, which could augment existing ECoG electrode arrays, to simultaneously measure intracranial pressure along with the neural signals.

This device could also be employed with brain implants that causes elevation in ICP due to tissue - implant interaction often leading to edema. This research explores the concept and feasibility for integrating the sensing component directly on to the ECoG electrode arrays.
ContributorsSampath Kumaran, Ranjani (Author) / Christen, Jennifer Blain (Thesis advisor) / Tillery, Stephen Helms (Committee member) / Greger, Bradley (Committee member) / Arizona State University (Publisher)
Created2015
Description
Cardiac tissue engineering has major applications in regenerative medicine, disease modeling and fundamental biological studies. Despite the significance, numerous questions still need to be explored to enhance the functionalities of the engineered tissue substitutes. In this study, three dimensional (3D) cardiac micro-tissues were developed through encapsulating co-culture of cardiomyocytes and

Cardiac tissue engineering has major applications in regenerative medicine, disease modeling and fundamental biological studies. Despite the significance, numerous questions still need to be explored to enhance the functionalities of the engineered tissue substitutes. In this study, three dimensional (3D) cardiac micro-tissues were developed through encapsulating co-culture of cardiomyocytes and cardiac fibroblasts, as the main cellular components of native myocardium, within photocrosslinkable gelatin-based hydrogels. Different co-culture ratios were assessed to optimize the functional properties of constructs. The geometry of the micro-tissues was precisely controlled using micro-patterning techniques in order to evaluate their role on synchronous contraction of the cells. Cardiomyocytes exhibited a native-like phenotype when co-cultured with cardiac fibroblasts as compared to the mono-culture condition. Particularly, elongated F-actin fibers with abundance of sarcomeric α-actinin and troponin-I were observed within all layers of the hydrogel constructs. Higher expressions of connexin-43 and integrin β1 indicated improved cell-cell and cell-matrix interactions. Amongst co-culture conditions, 2:1 (cardiomyocytes: cardiac fibroblasts) ratio exhibited enhanced functionalities, whereas decreasing the construct size adversely affected the synchronous contraction of the cells. Therefore, this study indicated that cell-cell ratio as well as the geometrical features of the micropatterned constructs are among crucial parameters, which need to be optimized in order to enhance the functionalities of engineered tissue substitutes and cardiac patches.
ContributorsSaini, Harpinder (Author) / Nikkhah, Mehdi (Thesis advisor) / Vernon, Brent (Committee member) / Towe, Bruce (Committee member) / Arizona State University (Publisher)
Created2015
154728-Thumbnail Image.png
Description
Several debilitating neurological disorders, such as Alzheimer's disease, stroke, and spinal cord injury, are characterized by the damage or loss of neuronal cell types in the central nervous system (CNS). Human neural progenitor cells (hNPCs) derived from human pluripotent stem cells (hPSCs) can proliferate extensively and differentiate into the various

Several debilitating neurological disorders, such as Alzheimer's disease, stroke, and spinal cord injury, are characterized by the damage or loss of neuronal cell types in the central nervous system (CNS). Human neural progenitor cells (hNPCs) derived from human pluripotent stem cells (hPSCs) can proliferate extensively and differentiate into the various neuronal subtypes and supporting cells that comprise the CNS. As such, hNPCs have tremendous potential for disease modeling, drug screening, and regenerative medicine applications. However, the use hNPCs for the study and treatment of neurological diseases requires the development of defined, robust, and scalable methods for their expansion and neuronal differentiation. To that end a rational design process was used to develop a vitronectin-derived peptide (VDP)-based substrate to support the growth and neuronal differentiation of hNPCs in conventional two-dimensional (2-D) culture and large-scale microcarrier (MC)-based suspension culture. Compared to hNPCs cultured on ECMP-based substrates, hNPCs grown on VDP-coated surfaces displayed similar morphologies, growth rates, and high expression levels of hNPC multipotency markers. Furthermore, VDP surfaces supported the directed differentiation of hNPCs to neurons at similar levels to cells differentiated on ECMP substrates. Here it has been demonstrated that VDP is a robust growth and differentiation matrix, as demonstrated by its ability to support the expansions and neuronal differentiation of hNPCs derived from three hESC (H9, HUES9, and HSF4) and one hiPSC (RiPSC) cell lines. Finally, it has been shown that VDP allows for the expansion or neuronal differentiation of hNPCs to quantities (>1010) necessary for drug screening or regenerative medicine purposes. In the future, the use of VDP as a defined culture substrate will significantly advance the clinical application of hNPCs and their derivatives as it will enable the large-scale expansion and neuronal differentiation of hNPCs in quantities necessary for disease modeling, drug screening, and regenerative medicine applications.
ContributorsVarun, Divya (Author) / Brafman, David (Thesis advisor) / Nikkhah, Mehdi (Committee member) / Stabenfeldt, Sarah (Committee member) / Arizona State University (Publisher)
Created2016
154575-Thumbnail Image.png
Description
The pathophysiology of neurodegenerative diseases, such as Alzheimer’s disease (AD), remain difficult to ascertain in part because animal models fail to fully recapitulate the complex pathophysiology of these diseases. In vitro models of neurodegenerative diseases generated with patient derived human induced pluripotent stem cells (hiPSCs) and human embryonic stem cells

The pathophysiology of neurodegenerative diseases, such as Alzheimer’s disease (AD), remain difficult to ascertain in part because animal models fail to fully recapitulate the complex pathophysiology of these diseases. In vitro models of neurodegenerative diseases generated with patient derived human induced pluripotent stem cells (hiPSCs) and human embryonic stem cells (hESCs) could provide new insight into disease mechanisms. Although protocols to differentiate hiPSCs and hESCs to neurons have been established, standard practice relies on two dimensional (2D) cell culture systems, which do not accurately mimic the complexity and architecture of the in vivo brain microenvironment.

I have developed protocols to generate 3D cultures of neurons from hiPSCs and hESCs, to provide more accurate models of AD. In the first protocol, hiPSC-derived neural progenitor cells (hNPCs) are plated in a suspension of Matrigel™ prior to terminal differentiation of neurons. In the second protocol, hiPSCs are forced into aggregates called embryoid bodies (EBs) in suspension culture and subsequently directed to the neural lineage through dual SMAD inhibition. Culture conditions are then changed to expand putative hNPC populations and finally differentiated to neuronal spheroids through activation of the tyrosine kinase pathway. The gene expression profiles of the 3D hiPSC-derived neural cultures were compared to fetal brain RNA. Our analysis has revealed that 3D neuronal cultures express high levels of mature pan-neuronal markers (e.g. MAP2, β3T) and neural transmitter subtype specific markers. The 3D neuronal spheroids also showed signs of neural patterning, similar to that observed during embryonic development. These 3D culture systems should provide a platform to probe disease mechanisms of AD and enable to generation of more advanced therapeutics.
ContributorsPetty, Francis (Author) / Brafman, David (Thesis advisor) / Stabenfeldt, Sarah (Committee member) / Nikkhah, Mehdi (Committee member) / Arizona State University (Publisher)
Created2016