Matching Items (2)
156230-Thumbnail Image.png
Description
Glucose sensors have had many paradigm shifts, beginning with using urine, to point of care blood, now being approved for implant. This review covers various aspects of the sensors, ranging from the types of surface chemistry, and electron transduction. All the way to the algorithms, and filters used to alter

Glucose sensors have had many paradigm shifts, beginning with using urine, to point of care blood, now being approved for implant. This review covers various aspects of the sensors, ranging from the types of surface chemistry, and electron transduction. All the way to the algorithms, and filters used to alter and understand the signal being transduced. Focus is given to Dr. Hellerâ’s work using redox mediators, as well as Dr. Sode in his advances for direct electron transfer. Simple process of designing sensors are described, as well as the possible errors that may come with glucose sensor use. Finally, a small window into the future trends of glucose sensors is described both from a device view point, as well as organic viewpoint. Using this history the initial point of care sensor for insulin published through LaBelle’s lab is reevaluated critically. In addition, the modeling of the possibility of continuously measuring insulin is researched. To better understand the design for a continuous glucose sensor, the basic kinetic model is set up, and ran through a design of experiments to then optimized what the binding kinetics for an ideal insulin molecular recognition element would be. In addition, the phenomena of two electrochemical impedance spectroscopy peaks is analyzed, and two theories are suggests, and demonstrated to a modest level.
ContributorsProbst, David L (Author) / LaBelle, Jeffery (Thesis advisor) / Caplan, Micheal (Committee member) / Cook, Curtiss (Committee member) / Arizona State University (Publisher)
Created2018
155424-Thumbnail Image.png
Description
The accurate and fast determination of organic air pollutants for many applications and studies is critical. Exposure to volatile organic compounds (VOCs) has become an important public health concern, which may induce a lot of health effects such as respiratory irritation, headaches and dizziness. In order to monitor the personal

The accurate and fast determination of organic air pollutants for many applications and studies is critical. Exposure to volatile organic compounds (VOCs) has become an important public health concern, which may induce a lot of health effects such as respiratory irritation, headaches and dizziness. In order to monitor the personal VOCs exposure level at point-of-care, a wearable real time monitor for VOCs detection is necessary. For it to be useful in real world application, it requires low cost, small size and weight, low power consumption, high sensitivity and selectivity.

To meet these requirements, a novel mobile device for personal VOCs exposure monitor has been developed. The key sensing element is a disposable molecularly imprinted polymer based quartz tuning fork resonator. The sensor and fabrication protocol are low cost, reproducible and stable. Characterization on the sensing material and device has been done. Comparisons with gold standards in the field such as GC-MS have been conducted. And the device’s functionality and capability have been validated in field tests, proving that it’s a great tool for VOCs monitoring under different scenarios.
ContributorsDeng, Yue, Ph.D (Author) / Forzani, Erica S (Thesis advisor) / Lind, Mary L (Committee member) / Mu, Bin (Committee member) / LaBelle, Jeffery (Committee member) / Emady, Heather (Committee member) / Arizona State University (Publisher)
Created2017