Description
Cellular heterogeneity is a key factor in various cellular processes as well as in disease development, especially associated with immune response and cancer progression. Cell-to-cell variability is considered to be one of the major obstacles in early detection and successful

Cellular heterogeneity is a key factor in various cellular processes as well as in disease development, especially associated with immune response and cancer progression. Cell-to-cell variability is considered to be one of the major obstacles in early detection and successful treatment of cancer. Most present technologies are based on bulk cell analysis, which results in averaging out the results acquired from a group of cells and hence missing important information about individual cells and their behavior. Understanding the cellular behavior at the single-cell level can help in obtaining a complete profile of the cell and to get a more in-depth knowledge of cellular processes. For example, measuring transmembrane fluxes oxygen can provide a direct readout of the cell metabolism.

The goal of this thesis is to design, optimize and implement a device that can measure the oxygen consumption rate (OCR) of live single cells. A microfluidic device has been designed with the ability to rapidly seal and unseal microchambers containing individual cells and an extracellular optical oxygen sensor for measuring the OCR of live single cells. The device consists of two parts, one with the sensor in microwells (top half) and the other with channels and cells trapped in Pachinko-type micro-traps (bottom half). When the two parts of the device are placed together the wells enclose each cell. Oil is flown in through the channels of the device to produce isolated and sealed microchamber around each cell. Different fluids can be flowed in and out of the device, alternating with oil, to rapidly switch between sealed and unsealed microenvironment around each cell. A fluorescent ratiometric dual pH and oxygen sensor is placed in each well. The thesis focuses on measuring changes in the oxygen consumption rate of each cell within a well. Live and dead cells are identified using a fluorescent live/dead cell assay. Finally, the technology is designed to be scalable for high-throughput applications by controlling the flow rate of the system and increasing the cell array density.
Reuse Permissions
  • Downloads
    pdf (1.8 MB)

    Details

    Title
    • Microfluidic platform for multi-parameter analysis of live single cells
    Contributors
    Date Created
    2014
    Resource Type
  • Text
  • Collections this item is in
    Note
    • Partial requirement for: M.S., Arizona State University, 2014
      Note type
      thesis
    • Includes bibliographical references (p. 53-54)
      Note type
      bibliography
    • Field of study: Engineering

    Citation and reuse

    Statement of Responsibility

    by Meryl Rodrigues

    Machine-readable links