  4. Pressure based spirometry: mobile spirometry using a pressure transducer
  5. Full metadata

Pressure based spirometry: mobile spirometry using a pressure transducer

Full metadata

Description

Spirometry is a type of pulmonary function test that measures the amount of air volume and the speed of air flow from a patient's breath in order to assess lung function. The goal of this project is to develop and validate a mobile spirometer technology based on a differential pressure sensor. The findings in this paper are used in a larger project that combines the features of a capnography device and a spirometer into a single mobile health unit known as the capno-spirometer. The following paper discusses the methods, experiments, and prototypes that were developed and tested in order to create a robust and accurate technology for all of the spirometry functions within the capno-spirometer. The differential pressure sensor is set up with one inlet measuring the pressure inside the spirometer tubing and the other inlet measuring the ambient pressure of the environment. The inlet measuring the inside of the tubing is very sensitive to its orientation and position with respect to the path of the air flow. It is found that taking a measurement from the center of the flow is 50% better than from the side wall. The sensor inlet is optimized at 37 mm from the mouthpiece inlet. The unit is calibrated by relating the maximum pressure sensor voltage signal to the peak expiratory flow rate (PEF) taken during a series of spirometry tests. In conclusion, this relationship is best represented as a quadratic function and a calibration equation is computed to provide a flow rate given a voltage change. The flow rates are used to calculate the four main spirometry parameters: PEF, FVC, FEV1, and FER. These methods are then referenced with the results from a commercial spirometer for validation. After validation, the pressure-based spirometry technology is proven to be both robust and accurate.

Date Created
2013
Contributors
  • Miller, Dylan (Author)
  • Forzani, Erica (Thesis advisor)
  • Trimble, Steve (Committee member)
  • Xian, Xiaojun (Committee member)
  • Arizona State University (Publisher)
Topical Subject
  • Mechanical Engineering
  • Health Sciences
  • Capnography
  • Capno-Spirometer
  • FVC
  • PEF
  • Pressure
  • Spirometry
  • Pressure transducers
Resource Type
Text
Genre
Masters Thesis
Academic theses
Extent
iv, 28 p. : ill. (some col.)
Language
eng
Copyright Statement
In Copyright
Reuse Permissions
All Rights Reserved
Primary Member of
ASU Electronic Theses and Dissertations
Peer-reviewed
No
Open Access
No
Handle
https://hdl.handle.net/2286/R.I.17939
Embargo Release Date
Wed, 04/29/2015 - 19:56
Statement of Responsibility
by Dylan Miller
Description Source
Viewed on Dec. 10, 2013
Level of coding
full
Note
Partial requirement for: M.S., Arizona State University, 2013
Note type
thesis
Includes bibliographical references (p. 28)
Note type
bibliography
Field of study: Mechanical engineering
System Created
  • 2013-07-12 06:24:43
System Modified
  • 2021-08-30 01:41:30
  •     
  • 1 year 4 months ago
Additional Formats
  • OAI Dublin Core
  • MODS XML

Quick actions

About this item

Overview
 Copy permalink

Explore this item

Explore Document

Share this content

Feedback

ASU University Technology Office Arizona State University.
KEEP

Contact Us

Repository Services
Home KEEP PRISM ASU Research Data Repository
Resources
Terms of Deposit Sharing Materials: ASU Digital Repository Guide Open Access at ASU

The ASU Library acknowledges the twenty-three Native Nations that have inhabited this land for centuries. Arizona State University's four campuses are located in the Salt River Valley on ancestral territories of Indigenous peoples, including the Akimel O’odham (Pima) and Pee Posh (Maricopa) Indian Communities, whose care and keeping of these lands allows us to be here today. ASU Library acknowledges the sovereignty of these nations and seeks to foster an environment of success and possibility for Native American students and patrons. We are advocates for the incorporation of Indigenous knowledge systems and research methodologies within contemporary library practice. ASU Library welcomes members of the Akimel O’odham and Pee Posh, and all Native nations to the Library.

Number one in the U.S. for innovation. ASU ahead of MIT and Stanford. - U.S. News and World Report, 8 years, 2016-2023
Maps and Locations Jobs Directory Contact ASU My ASU
Copyright and Trademark Accessibility Privacy Terms of Use Emergency COVID-19 Information