136597-Thumbnail Image.png
Description
In order to address infant respiratory distress syndrome, this study attempts to develop and characterize a textile strain gauge fabricated with stainless steel, wool, elastic, and tencel. Faire Isle knitted patterns are investigated in order to create channels of conductivity

In order to address infant respiratory distress syndrome, this study attempts to develop and characterize a textile strain gauge fabricated with stainless steel, wool, elastic, and tencel. Faire Isle knitted patterns are investigated in order to create channels of conductivity for a linear sensor. The effect linear yarn density on linearity and sensitivity and hysteresis of the sensors is also investigated for sensor optimization. It was found that there was a significant difference between the patterned and non-patterned samples. The patterned sensors were found to have a lower range of resistance than the non-patterned sensors and a smaller average standard of deviation between measurements. The 7 tension, lower linear yarn density, elastic patterned sample was the only sample to not exhibit hysteresis after three trials as well as have a linear range from 11.5cm to 13cm where the sensor behaves in accordance with a linear transfer function.
2.13 MB application/pdf

Download restricted. Please sign in.
Restrictions Statement

Barrett Honors College theses and creative projects are restricted to ASU community members.

Details

Title
  • Textile Strain Gauge for Respiratory Monitoring in Infants
Contributors
Date Created
2015-05
Resource Type
  • Text
  • Machine-readable links